LIDT Testing Example (1064nm Pulse Laser)
Laser Induced Damage Threshold (LIDT) Measurement Record
Sample / Product: BP1064-10
Laser Parameters: 1064 nm, 65 ns, Repetition Rate: 10 Hz, Focused Spot: d = 0.5 mm
| No. | Current I (A) | Energy Q (mJ) | Damage Threshold (J/cm²) | Remarks | Test Result |
|---|---|---|---|---|---|
| BP1064-10 | - | - | - | - | Measured damage threshold is approx. 4.8 J/cm². |
| 1 | 85 | 15.1 | 7.70 | ||
| 2 | 85 | 15.1 | 7.70 | ||
| 3 | 82 | 6.6 | 3.36 | ||
| 4 | 83 | 9.4 | 4.79 | ||
| 5 | 85 | 15.1 | 7.70 |